After you have optimized the beamline, your
next step is to make sure that the goniometer is aligned to the x-ray beam. The
goniometer alignment procedure requires that a CAT supplied phosphor ball be
mounted and optically aligned. It is very important that you align the phosphor
correctly – if you do not align the phosphor, then your ability to set
the goniometer vertical position correctly will be impossible. You will also
need to open the goniometer gui from the main sergui
setup file menu. The goniometer gui provides the user
with the capability to make small relative moves of the goniometer.

Once the phosphor is aligned
you will need to change the camera view to the top camera at a 1x zoom and open
the timing shutter. On 22ID, you should set the beamline slits to 20 by 20
micrometers. On both beamlines, you will need to change the attenuation and sample
lighting so that you can see a well defined small image on the phosphor ball,
as shown below. Now you will need to use the goniometer gui
to move the phosphor image onto the red cross hair, which is the goni center of rotation.

The phosphor image in the previous image is
off in the horizontal so a small horizontal adjustment is made in the goni gui, resulting in the following
image.

You may need to
experiment with the goni gui
to determine which direction is which in your image display. The goni gui validates the entry so
that a user can not make a mistake and try and move the goni
a mile or so.
Designed and maintained by the SER-CAT
Webmaster.
We welcome your questions and comments. Send
e-mail to: SER-CATwebmaster@anl.gov