The optimization of the
SER-CAT undulator (22ID) and bending magnet beamlines is performed by using the
SERGUI hutch page (shown below).

Wavelength
Section
The wavelength
section provides feedback (in blue) on the current x-ray energy, wavelength,
and the undulator Gap Energy (ID only). The energy and wavelength also have
desired fields, which are used to change the beamline energy. The energy
(wavelength) of the delivered x-rays can be changed by entering a value in the
desired field and hitting a carriage return. A large number of motors may be
required to move, so be patient and let the system move to the new energy
(wavelength).
22ID
(1) The value of the gap energy for normal operation
(below 13.0 keV) should be ~ 140 eV larger then the beamline energy. This
offset is inherent in all APS undulators and reflects the errors in calibrating
the position (gap) of mechanical jaws of the undulator to delivered x-ray
spectrum.
(2) If the beamline energy is larger than 13.0 keV, the
system will automatically move the undulator to use the third harmonic and the
Gap Energy will be 1/3 of the beamline energy.
(3) If the energy is moved below 8.0 keV, the vertical
focusing mirror will change lanes, moving from a palladium coating to a bare
glass coating. This is done to improve throughput as well as reduce any
harmonic component of the delivered x-ray beam. Conversely, if the energy
change is from low to high energy and the new energy is above 8.0 keV, then the
mirror coating will be changed back to palladium. The mirror lane change motion
is done slowly, so this type of energy change will take longer then normal.
22BM
When
the energy is moved more then 1,000 eV the sagittal focus motor needs to be
reset. The setup focus gui can be opened from the main sergui setup pull down
menu.

After the energy
change has completed, push the reset button. The terminal window that you
started sergui in will report back the status of the reset. This process can
take a few minutes depending on what the original value of the focus.
IMPORTANT: The beamline should be optimized
after any large energy change.
Optimize Section
The optimize
section provides the user the capability of optimizing 3 beamline parameter:
(1) Monochromator Tune
(2) Vertical Beam Position
(3) Horizontal Beam Position
Monochromator Tune: The beamline energy is selected by use of a double crystal monochromator
(cryogenically cooled Si (220) on 22ID and water cooled Si(111) on 22BM). The
optimize tune button will perform a rocking curve of the second crystal; in which
the Bragg angle of the second crystal is “rocked” about its current
position to maximize the delivered flux. If the “rocking curve”
does not measure the peak and 2 half heights, the system will automatically
repeat the scan. This procedure does
not expose the sample to x-rays as the detector is upstream of the experimental
timing shutter. The system will automatically perform a fit to the
measured data and move the monochromator tune to the optimal value.
Vertical
and Horizontal Beam Position: Once the
beamline energy has been changed, the delivered beam position at the sample
position could have also moved. The vertical and horizontal beam optimizations
will change a number of beamline conditions to perform these scans:
(1)
The attenuation
will be set to 30%;
(2)
The sample z
position will be moved 300 micrometers out of the beam to minimize any x-ray
exposure to a mounted sample;
(3)
The beam line
slits will be set such that the scanning direction is small (50 micrometers)
and the off axis is large (1 mm);
(4)
Open the
experimental timing shutter.
After the beamline
has been properly set, the system will scan the slits to measure either the
vertical or horizontal beam profile. The scan may be repeated if the peak and 2
half-heights are not recorded. The system will perform a fit to the measured
data and move the slits to the center of the x-ray beam. After a successful
scan, the system will return all of beamline components back to their pre scan
positions.
NOTE: The sample
position will track any change in beam position, as identified by the
difference in the old and new positions for the vertical and horizontal slits. It is always a good idea to mount a phosphor and follow the
procedure in OPTIMIZING SAMPLE ALIGNMNET to ensure that the delivered x-ray
beam is properly aligned.
The feedback
button has not been activated and is not functional.
Shutter
Section
The shutter section allows the user to open
or close the experimental timing shutter.
Attenuation
Section
This section
identifies if any filters are in the beam by identifying the current value of
the transmission (which will turn red if not at 100%). The desired field allows
a user to insert filters into the beam as a percent transmission. The filters
are discrete foils, and as such, not all desired transmission values can be
obtained. The system will insert the filter combination which will come closest
to the desired transmission.
Slit Section
The current values
of the vertical and horizontal slits are shown in here in blue. The desired
field allows the user to change the slit size by entering a new value and
hitting carriage return.
View Section
A live streaming
video feed from a camera in the experimental hutch is shown in this section.
The controls are active and the camera can be moved and have its zoom modified
from outside of the hutch
Designed and maintained by the SER-CAT
Webmaster.
We welcome your questions and comments. Send
e-mail to: SER-CATwebmaster@anl.gov